Home > Products > Ultrasonic Flow Meter > Portable Ultrasonic Flow Meter > Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type
Ultrasonic Flow Meter Portable Type

Ultrasonic Flow Meter Portable Type

The portable ultrasonic flowmeter has a built-in printer and SD card memory, which can print the set measurement value in real-time or at a regular time.

It is widely used in tap water, heating, water conservancy, metallurgy, chemical industry, machinery, energy, and other industries, and can be used for production monitoring, flow comparison, temporary detection, flow patrol inspection, water balance commissioning, heat network balance commissioning, energy conservation monitoring. It is a necessary T instrument for flow and heat detection.

Product Features:

Measuring liquid: water, seawater, alkali, and other single fluids, with turbidity less than 10000ppm and particle size less than 1mm;
Pipe material: steel, cast iron, PVC pipe, and other full pipes that can be penetrated by ultrasonic wave, with dense texture, no serious corrosion, and scaling, and a thin layer of dense lining is allowed;
Pipelining material: rubber, or unlined pipe diameter range: DN15mm~1200mm;
Length of straight pipe section: upstream>10D, downstream>5D (D: pipe diameter);
Display mode: double line 20 character LCD display;
Ambient temperature/humidity: host - 20 ℃~+50 ℃; Sensor - 20 ℃~+90 ℃ (normal temperature), - 20~+160 ℃ (high temperature).

Technical parameter:

Category

Performance & parameter

Host

Principle

Ultrasonic TDOA principle

Accuracy

Flow: better than ± 1%

Display

Connectable 2 × 10 Backlit Chinese characters or 2 × 20 character Western language LCD, supporting Chinese, English, Italian and French

Signal output

One 4-20mA current output, impedance 0~1K, accuracy 0.1% (optional)

One OCT pulse output (pulse width 6~1000ms, default 200ms)

One relay output

Signal input

3-circuit 4-20mA current input, accuracy 0.1%, can collect temperature, pressure, liquid level and other signals

Three wire PT100 platinum resistor can be connected to realize heat measurement

Data interface

Isolated RS485 serial interface, the flowmeter can be upgraded through PC, and support MODBUS and other protocols

Data record

Thermal printer, external SD card, capacity up to 2G

Pipeline condition

Pipe material

All PVC, aluminum, glass fiber reinforced plastic and other pipes are allowed to be lined

Inner diameter of pipe

15-6000mm

Straight pipe section

U10/D5

Measuring medium

Type

Water, sea water, industrial sewage, acid liquor, alcohol, beer, various oils and other homogeneous liquids that can conduct ultrasonic waves

Temperature

Temperature: - 30~90/160

Corrosivity

10000ppm and small bubble content

Flow speed

0~±10m/s

Work environment

Temperature

Host: - 20~60 ; Flow sensor: - 30~160

Humidity

Host: 85% RH; Protection class of sensor P67

Power Supply

Eight built-in 1.2V rechargeable N-MH batteries with a capacity of 2000mAH can work for 20 hours each time they are fully charged. The optional 4-20mA output signal can work for 8 hours continuously. AC90~260V power adapter for continuous measurement

Power consumption

1.5w